patient radiation protection

before exposure

customize prescribing dental images upon need for every patient

using proper equipment to decrease radiation

Total filtration

collimation

position indication device "PID"

during exposure

thyroid collar

lead apron

digital sensors or fast films

beam alignment devices

film holding device

exposure factor selection

KV= 60-80, Milliamperage=6-8 and shortest exposure time

proper technique

retake of any image must be avoided at all times!

after exposure

proper receptor handling

proper film processing/image retrieval

radiation monitoring badge

- recommended for both intra & extra- oral - effective lead shield should be 0.25mm at least or equivalent

the dental radiographer must avoid the primary beam

distance & position recommendation

stand at least 6 feet (2 meters) away from x-ray, or use a protective barrier

position 90-135 degree angle to the beam

shielding recommendation

protective barriers

minimum aluminum filtration required for machine operating

at or below 70KV = 1.5 mm Al

above 70KV = 2.5 mm Al

most effective method of reducing a patient's radiation exposure

if not available use fast films; f-speed or insight

operator protection

protector guideline

radiation monitoring

equipment monitoring for leakage

personal monitoring

recommended for intra-oral RGs only